HPS masthead
What's New?
. Fukushima Decontamination Report
. CRCPD & CDC Grants for Volunteer Corps
. America's Nuclear Future
. February Newsletter
. Boice Nominated President of NCRP
. February Journal
. February ORS
. Schauer Given the Butterfly Award from Image Gently
. Kase President's Report to IRPA
. IRPA13 Accepting Posters
Upcoming Events
. HPS Midyear - Issues in Waste Management
5-8 February 2012
Dallas, Texas
. NRC Regulatory Information Conference
NRC Regulatory Information Conference
13-15 March 2012
Rockville, Maryland
. NCRP Annual Meeting
12-13 March 2012
Washington, DC
. James E. Turner Memorial Symposium
Call for Abstracts
18-19 April 2012
Oak Ridge Associated Universities, Pollard Auditorium, Oak Ridge, Tennessee
. IRPA13
13-18 May 2012
Glasgow, Scotland
. Canadian Radiation Protection Association (CRPA) Annual Meeting
27-30 May 2012
Halifax, Nova Scotia
. ACS Undergrad Summer Schools
10 June- 20 July 2012
. Nuclear Regulatory Commission Meeting Webcasts
February 2012
Bethesda, Maryland
09 February 2012

Answer to Question #1238 Submitted to "Ask the Experts"

Category: Instrumentation and Measurements — Instrument Calibration (IC)

The following question was answered by an expert in the appropriate field:

Q
A dose calibrator linearity test can be done using the shield method (CaliCheck kit). I found that the results are frequently not reliable: 5% criteria was exceeded even if it was done in 3 minutes and the dose calibrator was functioning. How can one reduce possible errors or variability?
A
I have this question submitted to a representative of a major dose calibrator manufacturer and I will update this answer when additional information is obtained. However, in that these instruments are most often well-type ionization chambers with sensitive electrometer circuits, they are limited in the sense that there is no gas multiplication when the gamma ray interacts with the chamber wall or directly with the gas. Thus to increase the measured current, especially with a shielded vial for 99Mo breakthrough or an activity linearity test, you might try using a large quantity of radioactivity. Do check with the manufacturer's instrument manual, or contact the manufacturer directly, as there may be a problem with the equipment. David J. Allard, CHP
Answer posted on 5 November 2001. The information and material posted on this website is intended as general reference information only. Specific facts and circumstances may alter the concepts and applications of materials and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice specific to whatever facts and circumstances are presented in any given situation. Answers are correct at the time they are posted on the Website. Be advised that over time, some requirements could change, new data could be made available, or Internet links could change. For answers that have been posted for several months or longer, please check the current status of the posted information prior to using the responses for specific applications.