HPS masthead
What's New?
. Fukushima Decontamination Report
. CRCPD & CDC Grants for Volunteer Corps
. America's Nuclear Future
. February Newsletter
. Boice Nominated President of NCRP
. February Journal
. February ORS
. Schauer Given the Butterfly Award from Image Gently
. Kase President's Report to IRPA
. IRPA13 Accepting Posters
Upcoming Events
. HPS Midyear - Issues in Waste Management
5-8 February 2012
Dallas, Texas
. NRC Regulatory Information Conference
NRC Regulatory Information Conference
13-15 March 2012
Rockville, Maryland
. NCRP Annual Meeting
12-13 March 2012
Washington, DC
. James E. Turner Memorial Symposium
Call for Abstracts
18-19 April 2012
Oak Ridge Associated Universities, Pollard Auditorium, Oak Ridge, Tennessee
. IRPA13
13-18 May 2012
Glasgow, Scotland
. Canadian Radiation Protection Association (CRPA) Annual Meeting
27-30 May 2012
Halifax, Nova Scotia
. ACS Undergrad Summer Schools
10 June- 20 July 2012
. Nuclear Regulatory Commission Meeting Webcasts
February 2012
Bethesda, Maryland
07 February 2012

Answer to Question #236 Submitted to "Ask the Experts"

Category: Doses and Dose Calculations — Conversion of survey meter readings to dose

The following question was answered by an expert in the appropriate field:

Q
I am using 32P mounted on glass microscope slides (irradiated apatite grains). I need to record dose in microSievert but my measurements are in cps. I have a Berthold LB 1210B survey geiger counter with a counting area of 100 cm2. For this instrument, the calibration factor for 32P is .026 Bq /cm /s. The source is a sheet about 1.5 cm2 and let's say that I am measuring at 1 cm. Can you give me a simple conversion formula? Thanks!
A

It appears that the calibration factor given is incorrect. It is therefore assumed that the calibration factor for 32P is closer to 0.5 c/s per Bq--which would be similar to other thin-window GM detectors. As such, please refer to my work site and "Ask the Experts" question #82 for brief discussions of some beta dose-rate conversion factors. The values discussed in the above links are for pancake detectors and assume uniform irradiation across a surface area equal to 15.5 cm2. The data can easily be scaled for other thin-window detectors and exposure areas as well. For an exposed area equal to 1.5 cm2, the number of counts observed per unit dose absorbed will be an order of magnitude less- or about 24,000 to 30,000 cpm per mGy/h (400 cps per mSv/h).

Robert L. Coleman Oak Ridge National Laboratory

The information and material posted on this website is intended as general reference information only. Specific facts and circumstances may alter the concepts and applications of materials and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice specific to whatever facts and circumstances are presented in any given situation. Answers are correct at the time they are posted on the Website. Be advised that over time, some requirements could change, new data could be made available, or Internet links could change. For answers that have been posted for several months or longer, please check the current status of the posted information prior to using the responses for specific applications.
image
image
Home Affiliates Ask the Experts Radiation Terms Employment Meetings