HPS masthead
What's New?
. Fukushima Decontamination Report
. CRCPD & CDC Grants for Volunteer Corps
. America's Nuclear Future
. February Newsletter
. Boice Nominated President of NCRP
. February Journal
. February ORS
. Schauer Given the Butterfly Award from Image Gently
. Kase President's Report to IRPA
. IRPA13 Accepting Posters
Upcoming Events
. HPS Midyear - Issues in Waste Management
5-8 February 2012
Dallas, Texas
. NRC Regulatory Information Conference
NRC Regulatory Information Conference
13-15 March 2012
Rockville, Maryland
. NCRP Annual Meeting
12-13 March 2012
Washington, DC
. James E. Turner Memorial Symposium
Call for Abstracts
18-19 April 2012
Oak Ridge Associated Universities, Pollard Auditorium, Oak Ridge, Tennessee
. IRPA13
13-18 May 2012
Glasgow, Scotland
. Canadian Radiation Protection Association (CRPA) Annual Meeting
27-30 May 2012
Halifax, Nova Scotia
. ACS Undergrad Summer Schools
10 June- 20 July 2012
. Nuclear Regulatory Commission Meeting Webcasts
February 2012
Bethesda, Maryland
09 February 2012

Answer to Question #4105 Submitted to "Ask the Experts"

Category: Medical and Dental Equipment/Shielding — Shielding

The following question was answered by an expert in the appropriate field:

Q

What are the factors involved in determining the radiation protection features of a room suitable for barium enema investigations? The only thing I know so far is appropriate lead shielding for things like the doors, etc.

A

Barium enema studies have two parts, fluoroscopy and radiography. A typical radiographic and fluoroscopic (R&F) room has an under-the-couch screening system with two x-ray tubes. One tube is mounted in the tilting table for under-the-couch fluoroscopy with the x-ray receptor above the table. The other tube is mounted on the ceiling and used for radiography with a table Bucky as well as a vertical Bucky. For the radiographic images, the system is operated remotely with the operator located behind a shielded screen. Typically, various other examinations may also be performed in this room. 

Determination of shielding in the walls, floor, ceiling, and doors of an R&F room depends in part upon the weekly workload (mA-min/week) and distribution of operating potentials (kVp) with the workload, since not all procedures are performed at the same kVp. The weekly workload is the time integral of current, mA-min. The magnitude of the workload determines the radiation dose generated by the tube. The kVp distribution of the workload determines the dose as well as the transmission of the beam through the shielded barrier, for example, higher kVp values will require more shielding.

In addition to the operation parameters of the equipment, shielding requirements will depend on the dose being deposited in adjacent areas. The dose in any occupied location will be the sum of the doses from the primary x-ray beam, scattered radiation, and leakage radiation. The scattered radiation is the radiation that scatters from the patient when the primary x-ray beam is incident on the patient. The leakage radiation is the radiation that leaks through the x-ray source head, but not through the portal through which the primary beam emanates. During the fluoroscopy part of the exam, the image intensifier is well shielded and significantly attenuates the primary beam. Thus, only the scattered and leakage radiation needs to be considered in room shielding. For radiography exams, exposures can be made with the tube pointed down to the table Bucky or film cassette holder, sideways to the upright or wall-mounted Bucky, cross table for film cassette exposures, etc. The primary beam as well as the scattered radiation and leakage radiation should be considered for radiographic procedures for each direction the tube will be pointed.

Finally, the occupancy of areas around the x-ray room and distances from the x-ray source(s) to these areas of interest are needed to evaluate the expected doses to both occupational radiation workers and nonoccupational workers or the public at these locations. The above information together with the dose limits can be used to determine the transmission of the various barriers. The thickness of shielding required can then be determined.

The National Council of Radiation Protection and Measurements (NCRP) has issued a new document, NCRP Report 147, "Structural Shielding Design for Medical X-Ray Imaging Facilities (2004)," which will replace Report 49, "Structural Shielding Design and Evaluation for Medical Use of X-rays and Gamma Rays up to 10 MeV." A good summary on diagnostic x-ray shielding can be found on a website by Robert L. Dixon, Wake Forest University. Another helpful reference is "Radiation Shielding for Diagnostic X-Rays: Report of a Joint BIR/IPEM Working Party," edited by D.G. Sutton and J.R. Williams and published by the British Institute of Radiology, London, England.

With regard to regulatory requirements, each state has special requirements for the use of x rays in the healing arts, including fluoroscopic and radiographic installations. Information on specific state radiation protection programs requirements can be found at the Conference of Radiation Control Program Directors website.


Nisy E. Ipe, PhD, CHP

Answer posted on 6 January 2005. The information and material posted on this website is intended as general reference information only. Specific facts and circumstances may alter the concepts and applications of materials and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice specific to whatever facts and circumstances are presented in any given situation. Answers are correct at the time they are posted on the Website. Be advised that over time, some requirements could change, new data could be made available, or Internet links could change. For answers that have been posted for several months or longer, please check the current status of the posted information prior to using the responses for specific applications.
image
image
Home Affiliates Ask the Experts Radiation Terms Employment Meetings