Answer to Question #5410 Submitted to "Ask the Experts"
Category: Instrumentation and Measurements — Instrument Calibration (IC)
The following question was answered by an expert in the appropriate field:
Q
Many new GM-based dose-rate instruments are available in
addition to the standard ion-chamber instruments for gamma personnel
dose-rate surveys in an operating power reactor. What instrument issues
must be addressed in order to justify the use of GM instruments in a
predominately 60Co/16N environment for personnel dose-rate surveys? Most GM instrument manuals do not address 16N energy response. Instruments would be calibrated to 137Cs.
Why have ion-chamber instruments traditionally been recommended for
personnel dose rates? No one here seems to know the origin of the
requirement.
A
In response to the first part of your question, I would judge that,
commonly, the major issues to be considered in justifying the use of
Geiger Mueller (GM) detectors in a field dominated by gamma radiations
from 60Co and/or 16N are energy dependence of
dose response and dose-rate range capability. GM detectors, as you are
aware, are event detectors, and the rate at which individual events
take place represents a count rate, which is converted to exposure rate
or dose rate through the calibration process. There is no inherent
property of the detector that establishes a valid correlation between
exposure rate or soft-tissue dose rate and output of the detector.
When a detector is calibrated with 137Cs at a photon energy
of 662 keV, the major interaction mode in the detector is Compton
scatter, pretty much regardless of the materials of which the detector
is fabricated. Most GM detectors are fabricated of materials that have
an atomic number, Z, higher than that of air or soft tissue. The walls
are often metallic, and the most common filling gas is argon (e.g.,
steel with Z=26; Z for argon is 18). The effective atomic numbers for
air and water are each close to 7. While the Compton interaction in the
GM detector is dominant at 662 keV, this is not necessarily so at
lower and higher energies. Thus, as energy decreases the reduced energy
along with the elevated Z-values start to favor increased interactions
by the photoelectric process, which shows a strong inverse dependence
on energy, the cross section decreasing at about the inverse cube of
photon energy, and a very strong Z dependence, increasing at about the 4th to 5th
power of Z. Thus as energies decrease below a few hundred keV, the GM
response (count rate per unit exposure rate or per unit soft-tissue
dose rate) increases, yielding an inherent excess reading, relative to
the actual exposure or dose rate.
On the high-energy end of the spectrum, another differential response
is seen that can be attributed to the dominance of the pair production
event at high photon energies and to its strong dependence also on
atomic number, the pair production cross section increasing with the
square of Z. This effect accounts for a common over-response of many
GM instruments used in 16N photon fields, weighted heavily by 6 MeV gamma rays.
Many manufacturers reduce the lower-energy over-response by adding
additional metal around the detector to attenuate some of the low-energy photons. This process of energy compensation is usually
reasonably effective and may produce a detector that yields exposure
rate readings within 20 percent of the true values from roughly 200 keV
through energies somewhat greater than the approximate 1.25 MeV of 60Co.
Responses of such compensated detectors will usually drop off rapidly
as energies go below about 100 keV. The detectors will still
over-respond at the high energies such as those from 16N, with excess response of 50 percent to 100 percent being rather common.
Because of the very large gas discharge that accompanies a GM pulse,
the time required to clear the detector of sufficient charge to allow
the voltage to recover such that a subsequent event may be recorded is
quite long. This results in innate limitations on how rapidly the
detector may record events and places restrictions on how high a dose
rate may be handled by a standard G-M detector. The maximum reliable
dose rate varies depending on the design characteristics of the
detector and the associated electronics. Some of the older GM
detectors had the unfortunate characteristic of reading zero when
placed in a saturating field in which the detector could not recover.
Newer detectors generally show full-scale readings in such instances.
Also, with technology available over the past decade or so,
manufacturers have succeeded in using electronic means to provide
deadtime compensation so that the ranges of some GM detectors can be
extended considerably—e.g., a detector previously capable of a maximum
reading of 50 mR h-1 may be able to read as high as a couple of R h-1.
Regarding your question about the often-expressed preference for
ionization chambers for dose-related measurements in photon fields,
this preference stems from the fact that a properly designed ionization
chamber exhibits a response (e.g., amperes per unit exposure rate or
per unit dose rate) that is directly related to the true exposure rate
or dose rate—that is, the rate of collection of charge is equal to the
rate of production of primary ionization in the detector volume. This
is quite unlike the GM detector in which every output pulse is the
same regardless of how much primary ionization was produced in the
detector volume.
The proper design of the ionization chamber for exposure or dose
measurements can be achieved to an acceptable degree by abiding by
requirements of the Bragg-Gray principle, which I shall not discuss in
detail here (you can find discussions of this in many references in
health physics and instrumentation). One requirement, however, is that
the detector wall thickness be at least as great as the range of the
most energetic charged particle produced within it. Manufacturers
generally try to select a kind of compromise thickness, sufficient to
meet the needs for relatively high-energy photons but not so thick as
to produce excessive attenuation at lower energies.
Ionization chamber wall materials used for exposure or dose
measurements are often of low atomic number, and air is often the fill
gas, so that the kinds of photon interactions that occur in the
detector are very similar to those that occur in air or soft tissue at
the same energies over a wide range of energies. Detector walls of
commercial ionization chambers typically have density thicknesses on
the order of a few hundred mg cm-2, sufficient to provide a
near equilibrium distribution of secondary electrons over a range of
photon energies from less than 100 keV through 60Co but never high enough to ensure equilibrium for the high-energy electrons from 16N photons. As it often turns out, however, in the power reactor environments in which 16N is a concern, the photons emitted by 16N
in the coolant or steam lines frequently have had the opportunity to
interact in the coolant, pipe walls, ambient air, and other materials,
in the process producing a distribution of secondary electrons that
reasonably simulate an equilibrium distribution so that an air
ionization chamber placed in the field will often provide readings that
are within about 20 percent of the true exposure-rate or dose-rate values.
I apologize for the lengthiness of this response, but you have asked
some good questions that I felt required due consideration. I hope this
satisfies your needs.
George Chabot, PhD, CHP
Answer posted on 19 April 2006. The information and material posted on this Web site is intended as general reference information only. Specific facts and circumstances may alter the concepts and applications of materials and information described herein. The information provided is not a substitute for professional advice and should not be relied upon in the absence of such professional advice specific to whatever facts and circumstances are presented in any given situation. Answers are correct at the time they are posted on the Web site. Be advised that over time, some requirements could change, new data could be made available, or Internet links could change. For answers that have been posted for several months or longer, please check the current status of the posted information prior to using the responses for specific applications.